Notes when using biochemical testing machine

Hotline

0967 101 307

News & Event

Notes when using biochemical testing machine

In case of abnormal test results, we should repeat the test a few times if the results are still not accurate, then re-check how to mix chemicals and try again with new chemicals. If this still happens, call an engineer.
1. When turned on and off:
- When the power goes out, you need to turn off the machine immediately, until starting to have power, so the power source is stable for 3 to 5 minutes before we turn on the machine. Should use a voltage stabilizer to have a stable voltage.
- Be careful not to turn on the device immediately after turning off the device, because inductive current may damage electronic circuits. It takes at least 30 seconds for the device to stabilize to turn on the device.

2. When getting tested:
- Before starting the test, you should turn on the biochemical testing machine and wait 5 to 10 minutes for the device to stabilize before starting to test.
- New chemicals should be removed from the cabinet should be left for 15 to 20 minutes at room temperature.
- For tests using endpoint method (ENDPOINT), it should incubate enough time and at the right temperature.
- Avoid wiping the conical end, the suction head of the machine and the test tube with a dirty cotton or cloth. Because of this, it is easy to miss cotton and foreign objects that can clog in the measuring chamber and pump line leading to inaccurate test results.

- If the machine has a slow suction phenomenon, abnormal test results may be due to the following reasons:
+ Sampling samples are in accordance with regulations
+ How to suck and mix chemicals and specimens that are not right
+ How to preserve and use chemicals is not good
+ There have been changes in the installation or the chamber, the pump cable is blocked or for some other reason

- When encountering a slow suction machine, it must be washed, washed with distilled water or diluted javen 10 times. The butterfly needle can then be threaded into the pump head and a 10ml cylinder with a strong pump of distilled water to push the object. If slow suction continues, call an engineer.

- In case of abnormal test results, we should repeat the test a few times if the results are still not accurate, then re-check how to mix chemicals and try again with new chemicals. If this still happens, call an engineer.
- For accurate measurement results, the test should be recalculated every week in accordance with the chemical usage guidelines.

3. How to preserve the machine:
- Before turning off the machine must wash a lot with distilled water, cover the pump head to avoid dirt before turning off the power switch.
- Should be placed in a cool, dry place, not wet, dusty. It is best to have a dehumidifier or air conditioner.
- Every day, even if you do not perform the test, you should turn on the device for 30 minutes to prevent the device from getting wet or moldy.